Dräger Babylog VN500 Instructions For Use Manual page 227

Infinity acute care system
Table of Contents

Advertisement

Operating data (cont'd)
Internal battery Ventilation Unit
Type
Fuse
Capacity
Voltage
Current
Time bridged if mains supply is not available
with new and fully charged internal battery
Charging
Charging time (to charge battery complete-
ly)
Charging power
Internal battery PS500
Type
Capacity
Voltage
Time bridged if mains supply is not available.
with new and fully charged internal battery
Charging
Charging time (to charge battery
completely)
Charging power
Gas supply
O
pressure
2
O
peak input flow
2
O
connection
2
Air pressure
Air peak input flow
Air connection
Dew point
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
NiMH battery,
sealed, maintenance-free
F15A 80V UL248
2.5 Ah
24 V
0 to 15 A
min. 30 minutes (typically 40 minutes)
min. 4 hours
max. 50 W
VRLA batteries,
maintenance-free
(VRLA = Valve Regulated Lead Acid)
14 Ah to 16 Ah
36 V (3 x 12 V)
100 minutes (AC load 155 W / 175 VA)
min. 8 hours
max. 180 W
2.7 to 6.0 bar (or 270 to 600 kPa or 39 to 87 psi)
130 L/min (at 2.8 bar (40.6 psi) input pressure)
180 L/min (at 4.0 bar (58.0 psi) input pressure)
depending on configuration: DIN, NIST, DISS, Air
Liquide
2.7 to 6.0 bar (or 270 to 600 kPa or 39 to 87 psi)
130 L/min (at 2.8 bar (40.6 psi) input pressure)
180 L/min (at 4.0 bar (58.0 psi) input pressure)
depending on configuration: DIN, NIST, DISS, Air
Liquide
5 °C (41 °F) below ambient temperature
Technical Data
227