Automatic Interruption Of Oxygen Enrichment; After Reconnection - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Operation
enrichment program is terminated by
Babylog VN500 if there is no disconnection after
the 180 seconds have elapsed.
After disconnection for suction, Babylog VN500 de-
livers a minimal flow for the duration of disconnec-
tion in order to detect automatically the end of the
disconnection phase. 120 seconds are available for
suctioning. In the header bar, the disconnection
phase with the remaining time available for suction
is displayed continuously in seconds (C). If suction
is ended and the system is reconnected within the
displayed time, Babylog VN500 terminates the dis-
connection phase.
Automatic interruption of oxygen
enrichment
If there is no reconnection when the available time
(120 seconds) has elapsed, the oxygenation proce-
dure is terminated. All alarms are immediately reac-
tivated. Babylog VN500 immediately continues
ventilating in the set ventilation mode.

After reconnection

After reconnection, Babylog VN500 continues ven-
tilating in the set ventilation mode, except that for
120 seconds a correspondingly increased
O
concentration will continue to be delivered for fi-
2
nal oxygen enrichment.
In the header bar, the remaining time available for
the final oxygen enrichment phase is displayed
continuously in seconds.
Aborting oxygen enrichment
Touch the O
suction button again.
2
Additional information
The Suction maneuver function can be configured
as a O
suction button in the main menu bar to en-
2
able direct access. See "Assigning functions to ad-
ditional buttons" on page 146.
91
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n