Service Dialog; Device Extensions - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Configuration
External screen
If a second screen is connected to Infinity C500,
the user has to make a setting indicating whether
the screen is analog or digital.
Prerequisite: The System (A) page is opened.
1 Touch the Interface (B) tab.
2 Touch the External display (C) tab.
System setup
3 Touch the Digital or Analog button.
4 Confirm with the rotary knob.
Additional information
The serial interface connectors are located on the
rear of Infinity C500.

Device extensions

Prerequisite: The System (A) page is opened.
Touch the Supply units (B) tab.

Service dialog

The service dialog is password-protected and re-
served for DrägerService or trained service person-
nel.
164
A
C
Gas supply unit function GS500
B
The gas supply unit function can be deactivated if
Babylog VN500 is equipped with a gas supply unit
that is currently not supposed to be used.
The gas supply unit is no longer available. In the
device check, the system does not display the test
step GS500 (gas supply unit).
Activating the gas supply unit function:
1 Touch the On (D) button.
Babylog VN500 displays in the message field (E)
that the device check has to be carried out.
2 Perform device check.
Additional information
Using the gas supply unit, see "Gas supply unit
GS500" on page 99.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
System setup
D
C
Touch the Off (C) button.
E
A
B