Dräger Babylog VN500 Instructions For Use Manual page 166

Infinity acute care system
Table of Contents

Advertisement

Alarm – Cause – Remedy
Alarm
Alarm
priority
!!
050
"Audio paused" key used
too often
!!
050
"Audio paused" overused
or stuck
!
060
Accessory ID detection
failed
!!
110
Air and O
cient
!!
110
Air pressure regulation
failed
!!
110
Air supply insufficient
166
Cause
The "Audio paused" key is
either faulty or was pressed
more than 80 times per
hour.
The "Audio paused" button
is either stuck or faulty or
was pressed for more than
6 seconds.
Accessory ID detection mal-
function.
supply insuffi-
Air and O
2
cient to deliver required flow
and pressure for HFO.
Internal Air supply pressure
too high.
Air pressure regulator fail-
ure.
Air supply insufficient to de-
liver required flow and pres-
sure for HFO.
Driver gas for active expira-
tion during HFO is O
Instructions for Use
Remedy
The function of the "Audio
paused" key is not available
while the defect exists.
If the defect cannot be rem-
edied, call DrägerService.
Ventilation functions are not
affected.
Do not press the "Audio
paused" button longer than
6 seconds.
If the error persists, call
DrägerService.
Ventilation can be started
without ID functions.
Call DrägerService.
supply insuffi-
Check connections to Air
2
and O
Make sure supply pressures
are greater than 3 bar
(43.5 psi).
Downgrade alarm priority by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
No medication nebulization
possible.
Call DrägerService.
Check connection to Air
supply.
Make sure that the supply
pressure is greater than
only.
3 bar (43.5 psi).
2
Acknowledge message by
touching "ALARM RESET"
and confirm.
Infinity Acute Care System – Babylog VN500 SW 2.n
supply.
2