Reprocessing methods

Machine cleaning and disinfection

Use a washer-disinfector in accordance with
EN ISO 15883, preferably with a cart for anesthesia
and respiration accessories, for automatic cleaning
and disinfection. Use mild alkaline or enzymatic
(with neutral pH) cleaning agents. The user must be
sure to observe the manufacturer's information on
the cleaning agent exactly.
Placing parts in washer-disinfector
Place parts in washer-disinfector. Observe in-
structions for use of washer-disinfector.
Position parts so that all interior spaces are
completely flushed (e.g., hoses, valve assem-
blies) and water can drain off freely. Ensure all
areas are cleaned.
Keep spring-loaded valves open (water trap,
pneumatic medication nebulizer) during clean-
ing and disinfection, e.g., with rubber ring.
Cleaning program
Select the suitable program (preferably the an-
esthetics program). Cleaning is carried out at
40 to 60 °C (104 to 140 °F) for at least
5 minutes.
Thermal disinfection
1 Thermal disinfection is carried out at 80 to 95 °C
(176 to 203 °F) and with corresponding contact
time.
2 Carry out final rinsing with deionized water.
After ending cleaning and disinfection program
1 Immediately remove parts from washer-disin-
fector.
2 Inspect parts for visible soiling. If necessary, re-
peat cycle or clean manually.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Cleaning, Disinfection and Sterilization
3 Allow parts to dry thoroughly.
Cleaning agent
The material compatibility and effectiveness for re-
usable Dräger accessories was tested with various
mildly alkaline and enzymatic cleaning agents, and
at 93 °C (199.4 °F) for 10 minutes.
The test showed good material compatibility and ef-
fectiveness for the following cleaning agents:
– Neodisher MediClean manufactured by
Dr. Weigert
The user must be sure to observe the manufactur-
er's information on the cleaning agent exactly.

Manual cleaning

If no washer-disinfector is available, clean parts
manually under running water with cleaning agents.
Cleaning agent
The material compatibility and effectiveness for re-
usable Dräger accessories was tested with these
cleaning agents:
– Neodisher LM manufactured by Dr. Weigert
(concentration: 3 %, application time:
30 minutes)
The user must be sure to observe the manufactur-
er's information on the cleaning agent exactly.
199