Pc-Mmv - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

PC-MMV

Pressure Control-Mandatory Minute Volume
Ventilation
Pressure-controlled ventilation to ensure minimum
minute ventilation
Paw
PEEP
Flow
MV
Set MV
Mandatory MV
No spontane-
ous breathing
Pressure-controlled ventilation with volume
guarantee
The tidal volume of the mandatory breaths is deter-
mined by the volume VT. The duration of the man-
datory breaths is determined by Ti. The pressure
rise is determined by the Slope or Insp. flow set-
ting. The maximum pressure that Babylog VN500
uses is set via the Pmax therapy control. If Pmax is
not linked to the alarm limit Paw high, the pressure
can be limited using Paw high.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Pressure support PS
fast
pressure rise
Ti
1
1
RR
Insp. flow
beginning
spontaneous breathing
slow
pressure rise
Trigger window
Spontaneously breathed MV
In this case, the maximum applied pressure is lim-
ited to 5 mbar (5 cmH
maximum pressure allowed is not enough to deliver
the set VT, Babylog VN500 generates an alarm.
MMV works similar to SIMV, however, the manda-
tory breaths are only provided if spontaneous
breathing is not sufficient and below the prescribed
minimum ventilation. Should spontaneous breath-
ing increase, fewer mandatory breaths will be pro-
vided. The minimum ventilation is determined by
the setting of the tidal volume VT and the respirato-
ry rate RR.
Pressure limitation
Pmax
t
t
t
t
t
sufficient
spontaneous breathing
O) below Paw high. If the
2
Description
241