Alarm
Alarm
priority
!!
140
Check settings
!!
252
Check ventilation settings
!!
252
Check ventilation settings
!!!
144
Clean CO
!!!
145
CO
measurement failed
2
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Cause
Loss of stored data was de-
tected.
Due to data loss, the device
uses previous settings.
While adjusting ventilation
settings or alarm limits, a
power interruption occurred.
Data loss.
cuvette
Cuvette or sensor window is
2
soiled, e.g. with deposits
due to medication nebuliza-
tion.
CO
CO
rect.
sensor faulty.
2
measurement incor-
2
Alarm – Cause – Remedy
Remedy
Check all settings and adjust
if necessary.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
Check all therapy settings
and adjust them if neces-
sary.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
The device may apply de-
fault settings. Check ventila-
tion settings and alarm lim-
its. Acknowledge message
by touching "ALARM RE-
SET" button and confirm
with rotary knob.
The device may apply de-
fault settings. Check ventila-
tion settings and alarm lim-
its. Acknowledge message
by touching "ALARM RE-
SET" button and confirm
with rotary knob.
Use clean cuvette and/or
clean CO
sensor.
2
Perform zero calibration.
Replace faulty CO
sensor.
2
Use external CO
monitor-
2
ing and deactivate integrat-
ed CO
monitoring.
2
Call DrägerService.
173