Dräger Babylog VN500 Instructions For Use Manual page 173

Infinity acute care system
Table of Contents

Advertisement

Alarm
Alarm
priority
!!
140
Check settings
!!
252
Check ventilation settings
!!
252
Check ventilation settings
!!!
144
Clean CO
!!!
145
CO
measurement failed
2
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Cause
Loss of stored data was de-
tected.
Due to data loss, the device
uses previous settings.
While adjusting ventilation
settings or alarm limits, a
power interruption occurred.
Data loss.
cuvette
Cuvette or sensor window is
2
soiled, e.g. with deposits
due to medication nebuliza-
tion.
CO
CO
rect.
sensor faulty.
2
measurement incor-
2
Alarm – Cause – Remedy
Remedy
Check all settings and adjust
if necessary.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
Check all therapy settings
and adjust them if neces-
sary.
Acknowledge message by
touching "ALARM RESET"
button and confirm with rota-
ry knob.
The device may apply de-
fault settings. Check ventila-
tion settings and alarm lim-
its. Acknowledge message
by touching "ALARM RE-
SET" button and confirm
with rotary knob.
The device may apply de-
fault settings. Check ventila-
tion settings and alarm lim-
its. Acknowledge message
by touching "ALARM RE-
SET" button and confirm
with rotary knob.
Use clean cuvette and/or
clean CO
sensor.
2
Perform zero calibration.
Replace faulty CO
sensor.
2
Use external CO
monitor-
2
ing and deactivate integrat-
ed CO
monitoring.
2
Call DrägerService.
173