Operation
As soon as the screen is completely dark
Disconnect the mains plug from the mains pow-
er socket.
Detach the gas supply connectors for com-
pressed air (Air) and oxygen (O
outlets of the central gas supply.
CAUTION
Disconnect the compressed gas hoses from the
central gas supply. Otherwise minute internal
leaks could contaminate the central gas supply
through the reverse flow of supply gases.
CAUTION
The compressed gas hoses should only be un-
screwed from the ventilation unit after the connec-
tors have been removed from the wall outlets of
the central gas supply. Otherwise the compressed
gas hoses are not depressurized and the user
may be injured.

Storing Babylog VN500

Switch Babylog VN500 to energy-saving mode if
stored for longer periods.
1 End operation. See "Ending operation"
on page 103.
2 Set the toggle switch on the left side of
Babylog VN500 to
switching off the device.
3 If the power supply unit PS500 is connected,
also set the toggle switch at the rear of PS500
to
(off).
4 Disconnect the mains plug from the mains pow-
er socket.
104
) from the wall
2
(off) immediately after
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
If Babylog VN500 cannot be switched off on
account of a device malfunction
1 Open the device flap on the left side of
Babylog VN500.
2 Set the toggle switch to
Once the toggle switch has been pressed and the
mains plug is disconnected, Babylog VN500 can-
not be switched on.
Placing back into operation
1 Insert the mains plug into the mains power
socket.
2 Open the device flap on the left side of
Babylog VN500.
3 Set the toggle switch to (on).
4 Switch on Babylog VN500.
Additional information
If the power supply unit PS500 is connected, see
"Power supply unit PS500" on page 51.
(off).