Monitoring

Activating neonatal flow monitoring
Reactivate neonatal flow monitoring after exchang-
ing the neonatal flow sensor or as soon as possible.
Prerequisite: The Neonatal flow (A) page is
opened.
Touch the On (D) button and confirm with the
rotary knob.
Neonatal flow monitoring is activated.

Deactivating or activating O

WARNING
If O
monitoring is deactivated, ensure that an
2
appropriate replacement monitoring function
is available immediately. Other settings may
jeopardize the patient.
O
monitoring can be replaced by an appropriate
2
replacement monitoring. Set the O
the replacement monitoring according to the set
value FiO
:
2
FiO
<60 Vol%
–> O
2
≥60 Vol%
FiO
–> O
2
Deactivating O
monitoring
2
1 Touch the Sensors/ Parameters... button in the
main menu bar.
2 Touch the FiO
(A) tab.
2
Sensors/Parameters
A
D
B
3 Touch the Off (B) button and confirm with the ro-
tary knob.
138
monitoring
2
alarm limits of
2
±4 Vol%
2
±6 Vol%
2
C
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Babylog VN500 displays the following information
in the (C) message field: External monitoring
must be used!
O
monitoring is deactivated. Babylog VN500 dis-
2
plays the
FiO
symbol in the header bar. The
2
measured values are no longer displayed. The cor-
responding alarm function is deactivated.
Activating O
monitoring
2
O
monitoring should be reactivated as soon as
2
possible.
Prerequisite: The FiO
2
Touch the On (D) button and confirm with the
rotary knob.
O
monitoring is activated.
2
(A) page is opened.