Setting ventilation

Overview

This chapter describes how to set ventilation
modes and basic settings as well as additional set-
tings for ventilation parameters.
For a detailed description of the ventilation modes
and ventilation parameters, see chapter "Descrip-
tion of the ventilation modes" on page 236 and "Ad-
ditional settings for ventilation" on page 249.
Opening the Ventilation settings dialog
window
The Ventilation settings dialog window can be
opened as follows:
Touch the Ventilation settings... button (A) in
the main menu bar.
Touch the
(B) button in the therapy bar.
Touch the displayed ventilation mode (C) in the
header bar.
C
Babylog VN500 opens the Ventilation settings di-
alog window.
74
The page for the active ventilation mode (D) with
the Basic settings (F) appears by default. The cor-
responding therapy controls (E) are displayed.
The tab for Additional settings (G) can be used to
supplement the active ventilation mode with addi-
tional settings.
Ventilation settings
D

Selecting ventilation modes

Prerequisite: The Basic settings (A) page is
opened.
Ventilation settings
B
1 2 3
A
The Ventilation settings dialog window contains
5 tabs for selecting the ventilation modes. 4 tabs
(B) have ventilation modes permanently assigned
to them. The fifth tab (C) can be used to select an-
other ventilation mode, which can be selected from
B
the available ventilation modes.
The following 4 ventilation modes are preset at the
factory:
– PC-CMV
– PC-AC
– PC-SIMV
– PC-PSV