Dräger Babylog VN500 Instructions For Use Manual page 64

Infinity acute care system
Table of Contents

Advertisement

Getting Started
Test steps in the device check
In the device check the following test steps are car-
ried out:
– Auxiliary acoustical alarm (test of auxiliary
alarm/power failure alarm)
– Breathing circuit connection (visual inspection
of breathing circuit)
– Inspect humidifier (visual inspection of breath-
ing gas humidifier)
– CO
sensor calibration
2
– Neonatal flow sensor: Calibration
– Neonatal flow sensor: Measurement
– Test lung connection
– Gas supply sensors: Calibration
– O
supply
2
– Air supply
– GS500 (gas supply unit) (if the gas supply unit
function is activated)
– Pressure sensor calibration valve
– Expiratory valve (expiratory valve check)
– Safety valve (safety function check)
– O
sensor: Calibration
2
– Medication nebulizer (medication nebulizer
control check)
Device check procedure
Babylog VN500 guides the user in the form of a
question/answer dialog through the respective test
step. The instruction field (E) displays the questions
or instructions how to carry out the test steps.
The questions must be answered by pressing the
Yes (F) or No (G) buttons.
The Next test (H) button can be used to skip the
test steps.
A test step is also skipped if the necessary prereq-
uisites have not been met.
64
Start/Standby
E
The test steps in the device check are displayed
with the following symbols:
Rotating symbol
Green dot
Red dot
Colorless dot
Repeating test steps in the device check
1 Touch the Repeat (I) button.
2 Confirm with the rotary knob.
All test steps that have not yet been carried out or
that were unsuccessful are repeated.
Aborting the device check
1 Touch the Cancel (J) button.
2 Confirm with the rotary knob.
The device check is also aborted when the Device
check page is exited. The device check can be
continued when the Device check page is called
again.
1 Touch the Repeat (I) button.
2 Confirm with the rotary knob.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
J
I
F
G
H
:
Active test step
:
Correct result
:
Incorrect result
:
Test step not carried out

Advertisement

Table of Contents
loading

Table of Contents