Configuration
Table of alarm limits
The following table lists the alarm limits with the set-
ting range and the factory-set start-up values
(Dräger default). The hospital-specific start-up val-
ues can be entered in the table.
Parameter
MV
MV
Paw
RR
Tapn
etCO
2
etCO
2
Additional information
The alarm limits for the minute volume are set as a
percentage share of the start-up value (VT x RR).
To configure RR, see "Configuring start-up settings
for the ventilation parameters" on page 153.
For an overview of device's internal alarm limits,
see chapter "Automatic alarm limits" on page 231.
150
Setting range
1 to 100 %
1 to 100 %
7 to 105 mbar
(7 to 105 cmH
O)
2
5 to 200/min, Off
5 to 60 seconds
1 to 98 mmHg
0 to 97 mmHg
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Factory-set
start-up value
(Dräger default)
(VT x RR) +50 %
(VT x RR) –20 %
30 mbar
(30 cmH
O)
2
50/min
15 seconds
60 mmHg
30 mmHg
Hospital-specific set