3 Press the rotary knob to confirm the value. The
color of the therapy control turns dark green.
The following chapters of the Instructions for Use
provide a simplified explanation of these steps:
"Use the rotary knob to set and confirm the value."
Exceeding the set limit of a ventilation
parameter
When a set limit of a parameter has been reached,
Babylog VN500 displays a message.
Press the rotary knob to exceed the set limit.
The set limit can be exceeded.
If the maximum set limit for a parameter has been
reached, e.g., depending on other parameters, it is
not possible to exceed the set limit.
Press the rotary knob. Babylog VN500 adopts
the maximum possible set value.

Direct setting of ventilation parameters

(QuickSet)

When a ventilation parameter is set directly, the
changes to a setting become immediately effective
for the patient. The user can immediately see the
effect the changed setting has on the patient. The
finally chosen setting does not have to be con-
firmed again.
Ventilation parameters can be set directly in all ven-
tilation modes and can be carried out in the dialog
window for the ventilation settings. Direct settings
are only possible in the therapy bar when the ther-
apy controls are not locked.
O
and Flow cannot be set directly.
2
Setting ventilation parameters directly
1 Touch the corresponding therapy control.
2 Press and hold the rotary knob for approx.
3 seconds.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
The therapy control changes to dark green with a
yellow edge. The direct setting function is now ac-
tive.
3 Press and hold the rotary knob and turn to set
the value.
The set value is immediately effective.
Exceeding the set limit of a parameter with
direct setting
When a set limit of a parameter has been reached,
Babylog VN500 displays a message.
4 Release rotary knob for a short moment.
5 Press the rotary knob again and turn it.
The set limit can be exceeded.

Linked setting of ventilation parameters

The linked setting is possible for PEEP/Pinsp and
for RR/Ti.
Linking PEEP/Pinsp
Ventilation settings
C
1 Touch the therapy control PEEP (A) or
Pinsp (B); the color turns to yellow.
The Link (C) button is displayed.
2 Touch the Link (C) button.
Operating Concept
D
B
A
35