Switching On O 2 Therapy; Setting Fio 2 And Flow For O; Therapy - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Switching on O

therapy

2
The O
therapy can only be switched on in standby
2
mode.
1 Touch the Start/ Standby... button in the main
menu bar.
Babylog VN500 opens the Start/Standby dialog
window. The Start/Standby (A) page appears by
default.
2 Touch the Standby (B) button and confirm with
the rotary knob.
Babylog VN500 is in standby mode.
Start/Standby
A
E
B
C
3 Touch the O
Therapy (C) button.
2
The message field (D) displays the information to
use specific masks for the O
4 Connecting a mask, hood, or nasal cannula for
O
therapy.
2
5 Touch the Start (E) button and confirm with the
rotary knob.
O
therapy is switched on. Babylog VN500 displays
2
the main screen with the therapy bar (F) for the
O
therapy. The header bar (G) displays the mes-
2
sage O
Therapy.
2
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
D
therapy.
2
G
F
During the O
therapy, the screen display on the
2
main screen cannot be customized.
Setting FiO
and flow for O
2
C
A B
1 Touch the corresponding therapy control in the
therapy bar:
– FiO
(A)
2
– Flow (B)
2 Set the value by turning the rotary knob and
push to confirm.
The FiO
concentration is represented graphically
2
(C).
Operation
1 2 3
therapy
2
1 2 3
D
101