WARNING
The nebulizer port (F) must be used for nebu-
lization only! Otherwise the proper function-
ing of the device may be disrupted and the pa-
tient endangered.
Connect the nebulizer hose (G) to the nebulizer
port (F).
F
G
Fill the medication nebulizer in accordance with
the corresponding Instructions for Use.
CAUTION
Check the correct functioning of the medication
nebulizer. Check whether aerosol is generated.
A medication nebulizer fault is not detected by
Babylog VN500.

Switching on medication nebulization

1 Touch the Special procedures... button in the
main menu bar.
2 Touch the Med. nebulization (A) tab.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Special procedures
A
B
3 Touch the button for the desired nebulization
time (B).
Nebulization can be set for 5, 10, 15, 30 minutes.
Deactivating neonatal flow monitoring
Special procedures
A
Touch the Off button (D) and confirm with the
rotary knob.
Removing the neonatal flow sensor from the
breathing circuit
Special procedures
A
Operation
C
D
E
95