Zero Calibration - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Performing a CO
2
Zero calibration is performed with a clean CO
sor which is removed from the cuvette. Zero cali-
bration is performed in ambient air.
CAUTION
The cuvette windows of the reusable cuvette have
different optical properties to the cuvette windows
of the disposable cuvette.
The cuvette type used must therefore be selected
on the Zero calib. on/off page. Otherwise the
zero point is shifted by up to 8 mmHg CO
Prerequisite: Babylog VN500 is switched on and at
least the three-minute warm-up phase for the
CO
sensor has elapsed. After 3 minutes, the mea-
2
sured values will be inside the specified tolerance
range.
Starting zero calibration
1 Touch the Sensors/ Parameters... button in the
main menu bar.
2 Touch the CO
(A) tab.
2
The Zero calib. on/off (B) page appears by default.
Sensors/Parameters
A
C
3 Touch the Start (C) button.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n

zero calibration

2
.
2
When requested by Babylog VN500:
sen-
4 Remove CO
5 Confirm with the rotary knob.
Babylog VN500 performs zero calibration and dis-
plays the message Calibration in progress.
If zero calibration was successful
After approx. 5 seconds, Babylog VN500 reports
Zero calibration successful.
Fit the CO
2
If zero calibration was not successful
Babylog VN500 reports Zero calibration can-
celed.
Repeat zero calibration.
B
If zero calibration is still impossible
1 Check whether the sensor is soiled and clean it
if necessary. If the sensor is defective, replace
the sensor.
2 Repeat zero calibration.
D
E
sensor (D) from the cuvette (E).
2
sensor (D) back on the cuvette (E).
Monitoring
133