Operation
During manual inspiration, the pressure of the man-
datory breath is determined by the PmanInsp ther-
apy control (C).
Set and confirm the relevant values using the
rotary knob.
Monitoring during NIV
Use additional monitoring, e.g., external SpO
if necessary.
A time-lag Tdisconnect between 0 and
60 seconds can be set for the lower alarm limit of
the airway pressure.
Additional information
"Setting the alarm limits" on page 114.
84
,
2
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n