Configuration
Selecting the factory settings
Touch the Dräger default (H) button and con-
firm with the rotary knob.
The Dräger default button also sets other start-up
settings on the Ventilation page and the Alarms
page to the factory settings.
Tables for start-up values
The following tables show the factory-set start-up
values (Dräger default) for VT, RR, Slope and
Flow trigger. The hospital-specific start-up values
can be entered in the table.
The following table applies to the selection of start-
up values depending on the patient category:
Factory-set start-up value
Patient
VT
cat.
(mL)
Neo.
5.0
Ped. pat.
50
The following table applies to the selection of start-
up values depending on the body weight according
to the Radford nomogram:
Factory-set start-up value
Weight
VT
(kg)
(mL)
0.5
3.0
5
36
15
110
154
RR
Slope
(/min)
(s)
60
0.1
29
0.2
RR
Slope
(/min)
(s)
100
0.05
32
0.2
26
0.2
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Hospital-specific set start-up value
Flow
VT
trigger
(L/min)
(mL)
0.3
1.0
Hospital-specific set start-up value
Flow
VT
trigger
(L/min)
(mL)
0.2
1.0
1.0
RR
Slope
(/min)
(s)
RR
Slope
(/min)
(s)
Flow
trigger
(L/min)
Flow
trigger
(L/min)