Connecting the electrical mains power

supply

If the power supply unit PS500 is connected, see
"Power supply unit PS500" on page 51.
Connecting the electrical mains power supply
to Babylog VN500
The mains power must conform with the voltage
range specified on the rating plate (100 V to 240 V,
50/60 Hz).
1 Plug the appliance socket (A) onto the appli-
ance connector (B).
2 Position the power cable (C) in the clamp (D).
Fit the clamp into the housing (E). Tighten the
screw (F) (stress relief).
F
C
D
E
Insert the mains plug into the mains power
socket.
The LED
on Infinity C500 lights up green.
50
Checking the toggle switch on Babylog VN500
CAUTION
Do not press the toggle switch during ventilation.
Prerequisite: The flap on the left-hand side of the
device is opened.
B
A
WARNING
Do not simultaneously touch the connectors
of the interfaces (H) and the patient. Danger of
electrical shock.
Electricity supply from the internal battery
Babylog VN500 is provided with an internal battery
to ensure that operation of the device can continue
for at least 30 minutes (provided that the battery is
fully charged and new) following failure of the
mains power supply.
CAUTION
The internal battery is intended exclusively for
emergency use and not for normal operation! It is
essential to operate the device with mains power
again as soon as possible after switching over to
operation with the internal battery.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
I
O
G
Check whether the toggle switch (G) is set to
(on).
If the toggle switch is set to
gle switch to (on).