Description

HFO-Sigh

During high-frequency oscillation, a sigh can be set
(RRsigh).
RRsigh determines how often per minute the sigh
is to occur. A conventional pressure-controlled
breath is applied with the pressure Psigh. The du-
ration of this breath is set with Tisigh. The high-fre-
quency oscillations are discontinued for at least
150 ms prior to the sigh and are resumed 250 ms
after the sigh. The high-frequency oscillations start
with an expiration.
Rise time and rise form of the sigh are determined
by the configuration Slope adjustment (Slope or
Insp. flow).
The sigh can also be triggered with the Man. in-
sp./hold function. The duration of the sigh is deter-
mined by touching and holding the Man. insp./hold
button.
254
Paw
Flow
Tisigh
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Psigh
MAPhf
t
t