Dräger Babylog VN500 Instructions For Use Manual page 270

Infinity acute care system
Table of Contents

Advertisement

Description
The intermittent gas delivery is done by the nebuliz-
er mixer valve (25). The nebulizer changeover
valve (26) closes the nebulizer gas outlet when the
nebulizer function is not switched on.
The nebulizer mixer valve, the nebulizer
changeover valve, the ejector changeover valve,
the nebulizer outlet and the two pressure regulators
form the medication nebulization/ejector drive
(I) assembly.
The CO
concentration of the breathing gas can be
2
measured using the CO
sensor (30). CO
is mea-
2
2
sured according to an optical measuring principle in
the mainstream.
Babylog VN500 works at an elevated positive pres-
sure of up to 10 bar (1000 kPa or 145 psi) without
restrictions.
270
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n