Description
– Apnea ventilation must have been active for at
least 2 minutes.
– The alarm message MV low is not active.
– One of the following conditions must additional-
ly be met:
– The ratio of MVspon to MV must be greater
than 25 % and the ratio of MVleak to MV is
less than 40 %.
Or
– 80 % of the mandatory breaths must be trig-
gered spontaneously.
If apnea reoccurs within 3 minutes following auto-
matic termination of apnea ventilation in the Ped.
pat. patient category, the Auto return from Apnea
Vent. function is disabled until apnea ventilation is
terminated manually or another ventilation mode is
selected.
For configuration of the Auto return from Apnea
Vent. function, see "Configuring general settings"
on page 158.