Ambient Conditions; Set Values - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Technical Data

Ambient conditions

During operation
Temperature
Atmospheric pressure
Rel. humidity
During storage and transportation
Temperature
Atmospheric pressure
Relative humidity
Depending on the accessories used, more
stringent ambient conditions can apply. Obser-
ve the corresponding Instructions for Use.

Set values

The required parameters can be adjusted with the
therapy controls of Babylog VN500 without any
loss of accuracy. Only closed loop therapy controls
(e.g. VT or O
concentration) comprise the accura-
2
cy of the corresponding measurement technique.
Respiratory rate RR
Inspiratory time Ti
Maximum inspiratory time for flow cycled breaths
Timax
Pediatric patients
Neonates
Tidal volume VT
Pediatric patients
Neonates
Tidal volume for pressure support VT
Pediatric patients
Neonates
216
10 to 40 °C (50 to 104 °F)
700 to 1060 hPa (10.2 to 15.3 psi)
10 to 90 %, without condensation
–20 to 60 °C (–4 to 140 °F) without PS500
–15 to 40 °C (5 to 104 °F) with PS500
500 to 1060 hPa (7.3 to 15.3 psi)
5 to 95 %, without condensation
0.5 to 150/min
0.1 to 3 s
0.1 to 4 s
0.1 to 1.5 s
0.02 to 0.3 L under BTPS ±5 % of set value, or
±10 mL, whichever is greater
0.002 to 0.1 L under BTPS ±8 % of set value, or
±0.5 mL, whichever is greater
0.02 to 0.3 L under BTPS ±5 % of set value, or
±10 mL, whichever is greater
0.002 to 0.1 L under BTPS ±8 % of set value, or
±0.5 mL, whichever is greater
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n