Preparation
Connecting the system cable to Infinity C500
1 Unscrew the cover from the socket (A).
2 Insert the system cable connector (B) into the
socket (A). Ensure that the connector is insert-
ed with the correct orientation.
3 Screw the cover back on.
Connecting the system cable to Babylog VN500
1 Open the flap on the left-hand side of
Babylog VN500.
2 Route the system cable between
Babylog VN500 and the handle.
D
3 Clip the protective sleeve (C) immediately after
the connector (E) onto the system cable (D).
Align the protective sleeve so that the slots of
the protective sleeve are facing downwards and
upwards.
42
A
B
F
E
C
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
4 Insert the system cable connector (E) into the
socket until the connector audibly clicks into
place.
5 Insert the protective sleeve (C) into the protec-
tive plate (F) at the same time.
6 Turn the protective sleeve (C) by approx. 90°
until the protective sleeve clicks into place. The
cable is secured.
7 Close the left-hand flap.
Disconnecting the system cable from
Babylog VN500
1 Push the locking mechanism on the connector
(E) backwards and pull out the connector.
2 Turn the protective sleeve (C) by approx. 90°
and pull it out of the protective plate (F).
Fixing the system cable in the clamp (G)
G
1 Open the clamp cover (H).
2 Place the system cable into the clamp. Keep the
cable length short between the clamp and
Babylog VN500.
3 Close the clamp cover (H) and engage. Ensure
that the cover engages securely.
Removing the system cable from the clamp
1 Open the clamp cover.
2 Remove the cable from the clamp.
3 Close the clamp cover and engage.
H