Deactivating or activating monitoring

Monitoring is activated at the factory. Each monitor-
ing function can be deactivated separately.

Deactivating or activating neonatal flow

monitoring

WARNING
If flow and volume monitoring is deactivated,
ensure that an appropriate replacement moni-
toring function is available immediately. Other
settings may jeopardize the patient.
WARNING
No apnea monitoring takes place when flow
monitoring is deactivated. Use an indepen-
dent apnea monitoring.
CAUTION
Patient-triggered ventilation is not possible when
flow monitoring is deactivated.
Neonatal flow monitoring can be deactivated, e.g.:
– If the neonatal flow sensor has failed but cannot
be replaced immediately
– If medication nebulization is being performed
– To permit ventilation in the event of major tube
leakage
When changing to the NIV application mode,
Babylog VN500 deactivates neonatal flow monitor-
ing.
Patient-triggered ventilation is not possible when
neonatal flow monitoring is deactivated. The venti-
lation functions and ventilation monitoring are limit-
ed without neonatal flow monitoring.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
If the calibration was unsuccessful, the previously
valid calibration remains effective.
sen-
2
Without the neonatal flow sensor, the minute vol-
ume is not monitored! Independent apnea monitor-
ing is recommended.
Deactivating neonatal flow monitoring
1 Touch the Sensors/ Parameters... button in the
main menu bar.
2 Touch the Neonatal flow (A) tab if the page is
not already preset.
Sensors/Parameters
A
3 Touch the Off (B) button and confirm with the ro-
tary knob.
Babylog VN500 displays the following information
in the (C) message field: External monitoring
must be used!
Neonatal flow monitoring is deactivated.
Babylog VN500 displays the
symbol in the header bar. The measured values are
no longer displayed. The alarm function is deacti-
vated.
C
B
D
Neonatal flow
Monitoring
137