Using The Medibus Protocol; Lan And Usb Interfaces Of Infinity C500 - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Using the MEDIBUS protocol

NOTE
All transferred data is for information only and
should not be used as basis for diagnostic or ther-
apeutic decisions.
MEDIBUS is a software protocol for the transfer of
data between Babylog VN500 and an external
medical or non-medical device (e.g., patient moni-
tors or computers for data management systems).
The combination of Babylog VN500, Infinity C500
and the external device must comply with the re-
quirements of Directives IEC/EN 60601-1-1 and
IEC/EN 60601-1-2.
Additional information
"MEDIBUS for Evita Infinity V500 and
Babylog VN500" (9039527)
"Dräger RS MEDIBUS Protocol Definition"
(9028258)
Connecting an external device for using
MEDIBUS
A
Connect an external device to the interface
COM 1, COM 2 or COM 3 (A) of
Infinity C500. Use MEDIBUS cable 8416326.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Configuring the interface
A description is given in chapter "Configure inter-
faces" on page 163.

LAN and USB interfaces of Infinity C500

Use of LAN interfaces (A) of Infinity C500 is only
permitted for service purposes.
The USB interfaces (B) should only be used for
connecting a USB storage media or a USB SIM
card reader.
WARNING
Do not simultaneously touch the connectors
of the interfaces and the patient. Danger of
electrical shock.
B
Preparation
A
B
43