Alarms

Overview

Alarms are issued acoustically and visually. The
alarm tone can be suppressed for 2 minutes.

Display of alarms

In the event of an alarm, the system displays the
relevant alarm message in the alarm message field
(A). If the parameter field (B) is configured to dis-
play an individual parameter, the parameter field
(B) of the parameter triggering the alarm flashes.
C
If the alarm message field (A) contains more alarms
than can be displayed, the More... (C) button ap-
pears in the header bar. Touching this button opens
the page containing all the active alarms.
112
A
1 2 3
B
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
The Alarms dialog window provides the following
functions for selection:
– Setting alarm limits
– Displaying current alarms
– Alarm history
– Alarm settings

Alarm priorities

Babylog VN500 assigns the appropriate priority to
each alarm message.
The background color of the alarm message field
indicates the priority of active alarm messages. The
parameter field of the parameter triggering the
alarm flashes in the color matching the alarm prior-
ity.
Warning
High priority alarm
message
Caution
Medium priority
alarm message
Note
Low priority alarm
message
High priority alarm messages that are no longer ac-
tive are displayed in the background color of the
alarm message field.
Babylog VN500 generates different alarm tone se-
quences to display alarms acoustically. The alarm
tone sequences can be configured, see "Selecting
alarm tone sequences" on page 151.
Red background
Yellow
background
Cyan background