Set values (cont'd)
Activation state of Apnea Ventilation
Activation state of the function Automatic return
from Apnea Ventilation
Tidal volume during Apnea Ventilation VTapn
Pediatric patients
Neonates
Respiratory rate during Apnea Ventilation RRapn
Inspiratory flow Flow
Pediatric patients
Neonates
Inspiratory pressure Pinsp
Inspiratory pressure limit Pmax
O
concentration FiO
2
2
Positive end-expiratory pressure PEEP or intermit-
Δ
tent PEEP
intPEEP
Trigger sensitivity Flow trigger
Pressure support Psupp
Rise time for pressure support Slope
Pediatric patients
Neonates
Airway Pressure Release Ventilation APRV
Inspiratory time Thigh
Expiratory time Tlow
Maximum time of low pressure level in
APRV/PEF Tlow max
Inspiratory pressure Phigh
Expiratory pressure Plow
Termination criterion (expiratory flow)
Exp. term.
Automatic Tube Compensation ATC
Inner tube diameter Tube Ø
Endotracheal tube ET
Pediatric patients
Neonates
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
on, off
on, off
0.02 to 0.3 L under BTPS ±5 % of set value, or
±10 mL, whichever is greater
0.002 to 0.1 L under BTPS ±8 % of set value, or
±0.5 mL, whichever is greater
2 to 150/min
2 to 30 L/min
2 to 30 L/min
1 to 80 mbar (or hPa or cmH
2 to 100 mbar (or hPa or cmH
21 to 100 Vol%
0 to 35 mbar (or hPa or cmH
0.2 to 5 L/min
0 to 80 mbar (or hPa or cmH
0 to 2 s
0 to 1.5 s
0.1 to 30 s
0.05 to 30 s
0.05 to 30 s
1 to 80 mbar (or hPa or cmH
0 to 35 mbar (or hPa or cmH
1 to 80 %
2 to 8 mm (0.08 to 0.31 inch)
2 to 5 mm (0.08 to 0.2 inch)
Technical Data
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
217