The current loop is frozen. The loops are drawn in
blue. After "freezing", a cursor (C) is displayed
which can be moved with the rotary knob. The re-
spective values are displayed (D).
Recording up to 10 loops of automatic or spon-
taneous breaths
1 Touch the Draw (E) button.
2 Set how many loops should be recorded with
the rotary knob and push to confirm.
The set number is displayed in the button.
The contrast with which loops are displayed de-
creases depending how far back the recording was
made.
Additional information
A grid only appears if loops are displayed in the
complete curve field.

Freeze waveforms

The Freeze waveforms function can be configured
as a button in the main menu bar to enable direct
access. See "Assigning functions to additional but-
tons" on page 146.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Touch the Freeze waveforms button in the
main menu bar.
The current curves are immediately frozen. The
cursor (A) displays the time of "freezing" and the
value at the cursor position.
To display a measured value at a certain moment in
time:
Position the cursor on the time with the rotary
knob.
The measured value or the measured value pair
are displayed above the curve.
Operation
A
87