Starting Niv Ventilation; Setting Ventilation Parameters For Niv - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

3 Touch the Tube/NIV (A) tab.
Start/Standby
A
B
4 Touch the NIV (B) button and confirm with the
rotary knob.
Babylog VN500 is in the NIV application mode.
Babylog VN500 displays the
header bar
In the Neo. patient category, flow monitoring is de-
activated.

Starting NIV ventilation

Prerequisite: The Start/Standby dialog window is
opened.
1 Touch the Start/Standby (A) tab.
Start/Standby
A
B
The Start (B) button remains preset for
15 seconds.
2 Confirm with the rotary knob.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Babylog VN500 starts the therapy with the set ven-
tilation parameters. The main screen for ventilation
is displayed.

Setting ventilation parameters for NIV

Therapy control Timax
The therapy control Timax (A) limits the maximum
duration of supported breaths (Pressure Support,
Volume Support, PPS) because the inspiratory ter-
mination criterion may be ineffective with very high
leakages.
symbol in the
Ventilation settings
Therapy controls TmanInsp and PmanInsp
Prerequisite: The Neo. patient category and the
SPN-CPAP ventilation mode are set.
Ventilation settings
During manual inspiration, the duration of the man-
datory breath is determined by the TmanInsp ther-
apy control (B).
Set the ventilation parameters as described
under "Setting ventilation parameters"
on page 75.
A
Set the value for Timax by turning the rotary
knob and push to confirm.
B
C
Operation
83