Displaying Information On Alarms; Displaying Current Alarms; Displaying The Cause And Remedy For An Alarm; Acknowledging An Alarm Message That Is No Longer Active - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Displaying information on alarms

Displaying current alarms

To display the current alarms, proceed as follows:
1 Touch the alarm message in the header bar.
Or
1 Touch the Alarms... button in the main menu
bar.
2 Touch the Current alarms (A) tab.
Alarms
A
All the current alarm messages are displayed chro-
nologically with the corresponding duration, priority
and alarm message text in the (B) list.
Displaying the cause and remedy for an
alarm
1 Touch the alarm message or select it in the (B)
list with the rotary knob.
2 Touch the (C) button.
This displays the cause and remedy for the alarm
message selected.
3 Eliminate the fault.
Additional information
For a list of causes and remedies, see chapter
"Alarm – Cause – Remedy" on page 165.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
C
B
Acknowledging an alarm message that
is no longer active
After the fault has been eliminated, the alarm tone
is silenced. Medium and low priority alarm messag-
es expire automatically. High priority alarm mes-
sages continue to be displayed even after the
cause of the alarm has been eliminated and need
to be acknowledged.
A
Touch the ALARM RESET (A) button in the
header bar.
Acknowledging all alarm messages that
are no longer active
Prerequisite: The Current alarms (A) page is
opened.
Alarms
A
1 Touch the Reset all (B) button.
2 Confirm with the rotary knob.
The acknowledgeable messages are deleted in the
header bar and in the list containing the current
alarms. However, Babylog VN500 records all alarm
messages in the alarm history.
Alarms
1 2 3
B
113