Configuring start-up settings for the
ventilation parameters
Prerequisite: The Ventilation (A) page is opened.
1 Touch the Start settings (B) tab.
System setup
A
C
D
E
Babylog VN500 displays the following pages for the
ventilation start-up settings:
– VT, RR, Trigger
– Pressures, O
, I:E
2
– Other settings
– ATC
The VT, RR, Trigger (C) page appears by default.
Setting start-up values for VT, RR, Slope and
Flow trigger
Depending on the patient category or the patient's
weight, these start-up values can be set:
– VT
– RR
– Slope
– Flow trigger
Setting start-up values depending on the
patient category
1 If not yet preset, touch the Patient (D) button
and confirm with the rotary knob.
Babylog VN500 displays the start-up values for the
different patient categories (E).
2 Touch the appropriate button (E).
3 Set the value by turning the rotary knob and
push to confirm.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
After the start of ventilation, Babylog VN500 begins
ventilation with the start-up values, dependent on
the patient category set on the Start/Standby
page.
Setting start-up values depending on the body
height/weight
Prerequisite: The VT, RR, Trigger (C) page is
opened.
1 Touch the Weight (F) button and confirm with
the rotary knob.
B
System setup
C
Babylog VN500 displays the start-up values for the
different body weights (G).
2 Touch the appropriate button (G).
3 Set the value by turning the rotary knob and
push to confirm.
After the start of ventilation, Babylog VN500 begins
ventilation with the start-up values, depending on
the body height set on the Start/Standby page and
the ideal body weight derived from that, or with the
set start-up body weight in the Neo. patient catego-
ry.
Configuration
F
G
H
153