Cleaning, Disinfection and Sterilization
1 Press and hold the locking lever (A) on the un-
derside of the inspiratory unit.
2 At the same time, turn the inspiratory unit (B)
approx. 20° in a counterclockwise direction.
3 Remove the inspiratory unit from the fitting.
Dismantling the inspiratory unit
1 Remove the diaphragm with adapter (C) from
the fitting of the inspiratory unit.
2 Do not dismantle the inspiratory unit any further.
Reprocessing the inspiratory unit
Reprocess the inspiratory unit and diaphragm in
accordance with the reprocessing list,
see page 202.
After cleaning
After cleaning, always sterilize the inspiratory
unit in hot steam at 134 °C (273.2 °F) to ensure
all remaining liquid is dried completely in the in-
terior areas.
198
A
B
C
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Dismantling and reprocessing
accessories
Dismantle and reprocess the breathing gas hu-
midifier, the Aeroneb Pro nebulizer and bacteri-
al filter in accordance with the corresponding In-
structions for Use.