Calibrating The Co 2 Sensor; Selecting The Cuvette Type; Sensor - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Calibrating the CO
Prerequisite: The CO
2
The CO
sensor is works-calibrated and can be
2
used on any ventilation unit.
The following cuvettes can be used:
– Reusable cuvettes
– Disposable cuvettes
The cuvette type used must be selected on the
Zero calib. on/off page.

Selecting the cuvette type

CAUTION
The cuvette windows of the reusable cuvette have
different optical properties to the cuvette windows
of the disposable cuvette.
The cuvette type used must therefore be selected
on the Zero calib. on/off page. Otherwise the
zero point is shifted by up to 8 mmHg CO
1 Touch the Sensors/ Parameters... button in the
main menu bar.
2 Touch the CO
(A) tab.
2
The Zero calib. on/off (B) page appears by default.
Sensors/Parameters
A
C
D
3 Touch the Reusable (C) or Disposable button
(D) and confirm with the rotary knob.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n

sensor

2
application is activated.
.
2
If the selected cuvette does not correspond to the
cuvette used, the alarm message Check cuvette is
displayed.
B
Monitoring
131