Assembly
Connecting the breathing hoses and
flow measuring lines
B
B
1 Connect the breathing hose (A) to the breathing
valve.
2 Connect the flow measuring lines (B) to the
nozzles of the flow sensor. When connecting
the flow measuring lines, pay attention to the
differing diameters of the hoses and nozzles
and connect them on the correct side.
44
A
B
A
3 Connect the flow measuring lines (B) to the
Oxylog 3000 plus. Correct alignment is
indicated by a notch on the connector, which
must point away from the breathing hose. If it is
incorrectly seated, incorrect values will be
measured.
4 Connect the breathing hose (A) to the gas outlet
on the Oxylog 3000 plus.
When connecting a breathing hose, check that
the hose setting in the settings window
corresponds to the connected hose.
Instructions for use Oxylog 3000 plus SW 1.n