Alarm – Cause – Remedy
The alarm messages are displayed in the message
field of the header bar in hierarchical order.
Different background colors indicate the priority
level of the alarm.
In the tables for Current alarms and Alarm
history, the priority of the alarm messages is also
indicated by exclamation marks.
Warning
!!!
Red
Caution
!!
Yellow
Note
!
Cyan
In order to classify the alarms within an alarm cate-
gory, internal priority numbers are given after the
exclamation marks in the table below. The most
critical alarm is awarded the number 255. The pri-
ority of the alarm decreases the lower the number
is.
If several alarms occur simultaneously, the mes-
sage field in the header bar first displays the two
most critical alarms.
In the following table, the alarm messages are list-
ed in alphabetical order. If an alarm occurs, the ta-
ble helps to identify causes and remedies. The dif-
ferent causes and remedies should be worked
through in the order listed until the alarm has been
resolved.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
High priority alarm
message
Medium priority
alarm message
Low priority alarm
message
Immediate action is necessary in order to avert an
imminent danger
Prompt action is necessary in order to avert a danger
Attention is necessary, but a delayed response is
sufficient
Alarm – Cause – Remedy
165