Importing And Exporting Configurations - Dräger Babylog VN500 Instructions For Use Manual

Infinity acute care system
Table of Contents

Advertisement

Importing and exporting configurations

Babylog VN500 can export the device configuration
to a USB storage media. The configuration saved
on the USB storage media can be imported to other
devices.
The following settings can be exported:
– Screen layout
– Alarms
– Ventilation
– Exchange intervals
– System
Only the values from the system setup are export-
ed.
Views are only exported if the view configured was
first saved on the Screen layout page.
Insert the USB storage media into a USB port
on Infinity C500.
1 Touch the System setup... button in the main
menu bar.
2 Touch the Config. exchange (A) tab.
System setup
A
B
Babylog VN500 displays the configuration on the
device (B) and the configuration available on the
USB storage media (C). The date of the import and
export is displayed.
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
C
D
E
Importing a configuration from a USB storage
media to the device
A configuration can only be imported in standby
mode.
1 Switch Babylog VN500 to standby mode.
2 Touch the
Import (D) button.
3 Confirm with the rotary knob.
If there is no valid configuration saved on the USB
storage media, the system issues a message.
After the import, Babylog VN500 is switched off
automatically.
4 Switch on Babylog VN500 again.
Babylog VN500 reports the completion of the con-
figuration with a low priority alarm.
5 Check the settings after the configuration.
Exporting a configuration from the device to a
USB storage media
1 Touch the
Export (E) button.
2 Confirm with the rotary knob.
If the USB storage media already contains a config-
uration, you are informed with a message stating
that this configuration will be overwritten.
Configuration
159

Advertisement

Table of Contents
loading