Displayed measured values
Accuracy does only apply for the measurement
range specified.
Airway pressure measurement
Plateau pressure
Positive end-expiratory pressure
Peak Inspiratory Pressure
Mean airway pressure
Minimal airway pressure
Lower pressure level in APRV
End-inspriatory pressure for mandatory bre-
aths
Upper pressure level in APRV
Pressure amplitude (peak-to-peak) in HFO
Intrinsic PEEP obtained during PEEPi maneu-
ver
Range
Accuracy
T0...90 (for Pmean)
O
measurement (inspiratory side)
2
Inspiratory O
concentration (in dry air)
2
Range
Accuracy
Drift of measurement accuracy
T0...90
Warm-up time
Flow measurement (proximal)
Minute volume measurement
Total minute volume
Mandatory minute volume
Spontaneous minute volume
Range
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
Pplat
PEEP
PIP
Pmean
Pmin
Plow
EIP
Phigh
Δ
Phf
PEEPi
–60 to 120 mbar (or hPa or cmH
±6 % of measured value, or ±0.5 mbar (or hPa or
cmH
O), whichever is greater
2
33 s for intubated patients and 20 s with NIV
FiO
2
18 to 100 Vol%
±3 Vol% at 20 °C (68 °F)
0.2 Vol% per month (the measured values are ba-
rometrically pressure compensated)
<30 s
max. 3 minutes, typ. 1 minute
MV
MVmand
MVspon
0 to 99 L/min
BTPS
Technical Data
O)
2
221