For the order numbers of the holder for the breath-
ing gas humidifier, see the list of accessories.

Attaching breathing hoses

WARNING
Do not use antistatic or conductive breathing
hoses. The use of these materials increases
the danger of electric shock to the patient and
of fire in an oxygen-enriched environment.
CAUTION
The sterile packaging of disposable articles
should only be opened immediately before use.
Otherwise there is a risk of infection.
1 Hang the hinged arm (A) on the lateral standard
rail of Babylog VN500 and tighten the screws.
Depending on the desired position of the device
in relation to the bed, the hinged arm can be fit-
ted to either side of Babylog VN500.
A
E
D
2 Connect breathing hoses to the inspiratory port
(B) and to the expiratory port (C).
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
CAUTION
Do not reverse the connections for inspiration (B)
and expiration (C). Humidification is ineffective if
the connections are reversed.
3 Turn the inspiratory port and the expiratory port
A water trap is required for the
Fisher & Paykel MR 850 breathing gas humidifier
depending on the breathing circuit used.
4 If a water trap is required, install the water trap
5 Connect the Y-piece (E) to the breathing hoses.
6 Insert the Y-piece or the breathing hoses in the
Using the Infinity ID breathing circuit
Babylog VN500 recognizes the use of an Infinity ID
breathing circuit. The message Infinity ID Breath-
ing Circuit detected is displayed in the header bar.
The following Infinity ID functions are supported:
– Detection of reversed hoses
– Detection of non-compliance with the patient
C
B
– Automatic configuration of breathing circuit and
These functions are only supported in standby
mode.
Setting the breathing circuit
Babylog VN500 supports the user in selecting the
breathing circuit on the page Start/Standby > Br.
circuit/ Humidifier.
in the direction of hoses.
(D) in a vertical position.
opening of the hinged arm.
category
humidifier
Fit the Infinity ID breathing hoses in standby
mode.
Set the breathing circuit according to the patient
category.
Preparation
47