Dräger Babylog VN500 Instructions For Use Manual page 155

Infinity acute care system
Table of Contents

Advertisement

Setting start-up values for pressures, FiO
I:E
Prerequisite: The Start settings (A) page is
opened.
1 Touch the Pressures, O
System setup
B
C
D
E
F
G
2 Touch the corresponding button for the
parameters:
– Pressures (C)
– APRV pressures (D)
– Insp. flow (E)
– FiO
(F)
2
– I:E (G)
3 Set the value by turning the rotary knob and
push to confirm.
Selecting the factory settings
Touch the Dräger default (H) button and con-
firm with the rotary knob.
The Dräger default button also sets other start-up
settings on the Ventilation page and the Alarms
page to the factory settings.
Parameter
Factory-set start-up value
PEEP
ΔPsupp
Pinsp
Pmax
Plow
Phigh
Instructions for Use Infinity Acute Care System – Babylog VN500 SW 2.n
and
2
, I:E (B) tab.
2
A
h
(Dräger default)
5 mbar (5 cmH
O)
2
0 mbar (0 cmH
O)
2
15 mbar (15 cmH
O)
2
40 mbar (40 cmH
O)
2
5 mbar (5 cmH
O)
2
15 mbar (15 cmH
O)
2
Hospital-specific set
start-up value
Configuration
155