Using the MEDIBUS or MEDIBUS.X protocol
NOTE
All transferred data is for information only and
must not be used as basis for diagnostic or
therapeutic decisions.
MEDIBUS and MEDIBUS.X are software protocols
for data transfer between Carina and other medical
devices (e.g., patient monitors) or other devices
(e.g., computers for data management systems).
Requirements for the combination of Carina and an
external device, see "Connection to other devices"
on page 9.
Observe software protocols
For MEDIBUS:
MEDIBUS for Carina
Dräger RS 232 MEDIBUS, Protocol
Definition
For MEDIBUS.X:
MEDIBUS.X, Rules and Standards for
Implementation
MEDIBUS.X, Profile Definition for
Data Communication V1.0
Instructions for use Carina SW 3.2n

Connecting an external device for

MEDIBUS or MEDIBUS.X

WARNING
Risk of electric shock
If the connectors of the interface and the
patient are touched simultaneously, there is a
risk of electric shock.
Do not simultaneously touch the connectors
of the interface and the patient.
The port
connecting devices with safety extra-low voltage
(SELV).
Use only "Babylink cable" 8306488 from Dräger.
9039037
9028258
9052607
9052608
Connect the external device to the port
Assembly and preparation
for RS232 is intended exclusively for
A
(A).
51