Testing The "Paw T1" Alarm.19; Testing The "Paw J" Alarm.19 - Dräger Oxylog 1000 Instructions For Use Manual

Emergency ventilator
Table of Contents

Advertisement

Checking readiness for operation
Testing the »Paw
f
«alarm
Testing the »Paw
« alarm and audible alarm muting
Testing the »Paw
alarm
1
Turn the »MV« knob to approx. 10 L/min.
2
Turn the
»Freq.«
knob to approx. 20 breaths per
minute.
3
Turn the
»Paw«
knob to 40 mbar.
4
Hold the test lung in the squeezed position and watch
the pressure gauge:
• Oxylog 1000 limits the airway pressure to approx.
40 mbar.
5
The
»Paw
J
T « indicator turns red, and the audible
alarm is sounded.
4
Release the test lung.
5
The
»Paw
J
T « indicator and audible alarm are
deactivated.
Testing the »Paw
« alarm and audible
alarm muting
With the existing setting
6
Remove the test lung from the ventilation valve:
7
The red
»Paw
J~
« indicator is displayed,
and the audible alarm is sounded.
8
Press the »<&« button.
• The
»Paw
J
~« remains red;
the audible alarm is muted for about 2 minutes.
9
The indicator
»<& «
turns yellow to indicate that the
alarm tone has been muted for about 2 minutes.
After about 2 minutes:
• the audible alarm is reactivated.
6
Refit the test lung:
7
The
»Paw
J~
« indicator and audible alarm are
deactivated.