Configuration
Displaying configuration and information
The actual screen display may differ in appearance
or configuration.
1 To switch the device ON briefly press the
key. The device performs a selftest and the user
is prompted to activate the configuration menu
or device check:
Press rotary knob for device check and
configuration.
2 Press the rotary knob to confirm, before the
progress bar is complete.
The start-up menu is then displayed:
3 Select Configuration and Information and
confirm.
The settings made in the Configuration window
are retained after the ventilator is switched OFF.
Configuration can be canceled by pressing the
Alarm reset key.
The following data can be displayed in the
Configuration and Information windows:
– Language
– Identification No. (Device ID)
– Total hours of operation (Working hours)
– Hours of operation since the last inspection and
maintenance (Hours since service time)
– Battery type and battery capacity.
96
Language
1 Press the key Settings
menu Configuration.
2 Select and activate the line Language.
3 Select the language and confirm.
The new language selected is effective
immediately.
Battery type
Press the key Settings
menu Information.
The performance data of the inserted battery
are displayed on the device.
Instructions for use Oxylog 3000 plus SW 1.n
to select the
to select the