Changed Parameter Syntax and Semantics

Table 13 Commands with Different Parameters or Syntax

Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
Changed Parameter Syntax and
Semantics
Table 13 details commands whose parameter syntax or semantics have
changed.
Command
SOUR:AM:FREQ/?
DISP:BRIG
SENS:CORR:COLL:ZERO?
SENS:POW:REF
SENS:POW:REF:STAT:RAT Accepts TOREF,0 or values for slot,channel,
SYST:DATe
GPIB Command Compatibility List
Change
This command does not accept the value
CW, instead use SOUR:AM:STAT ON|OFF
to switch from and to CW mode.
The commands accepts floating point val-
ues.
This command now supports integers be-
tween 1 and 100, instead of float values be-
tween 1 and 0.
This command returns the last zero state,
instead of the last remote zero state.
Accepts TOMODule and TOREF for the first
parameter, instead of accepting TOA|TOB
as the HP 8153A does. The numbers 0|1|2
cannot be used, only the strings above.
instead of accepting TOA|TOB as the
HP 8153A does. The numbers have a differ-
ent meaning.
SYST:DATe from HP 8167/8 is not support-
ed, but SYST:DATe from HP 8153 is sup-
ported.
Affects
8153
8153
8153
8153
8153
8167/8
251