Measurement Operations & Settings
command:
syntax:
description:
N O T E
parameters:
response:
example:
affects:
command:
syntax:
description:
N O T E
parameters:
response:
example:
affects:
170
:ROUTe[n][:CHANnel[m]]:CONFig?
:ROUTe[n]:[CHANnel[m]]:CONFig?
Queries the switch configuration of the instrument. For each channel, the minimum and
maximum channel number of each port is given.
none
<j>,<k>;<l>,<m> as text string where:
<j> is the first port character on the left
<k> is the last port character on the left
<l> is the minmimum port number on the right
<m> is the maximum port number on the right
rout2:conf? → A,B;1,2 (2 left and 2 right ports)
All switch modules
:ROUTe[n][:CHANnel[m]]:CONFig:ROUTe?
:ROUTe[n]:[CHANnel[m]]:CONFig:ROUTe?
Queries the allowed switch routes of an instrument.
none
[A.....Z],[1.....n];[A.....Z],[1.....n].[A.....Z],[1.....n] as a text string.
"," separates input and output ports of a single connection.
";" separates parallel connections
"." separates possible switch states
rout2:conf:rout? → A,1;B,2.A,2;B,1
2x2 x-over switcch:
state 1: When A to 1 then B to 2nd connection (straight)
state 2: When A to 2 then B to 1st connection (cross-over)