Equally Spaced Datapoints; Figure 18 Equally Spaced Datapoints - Agilent Technologies 8163A Programming Manual

Lightwave multimeter; lightwave measurement system; lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

The Agilent 816x VXIplug&play Instrument Driver
Lambda Scan Applications

Equally Spaced Datapoints

A linear interpolation is performed on all wavelength and power data for
the Lambda Scan Application and is optional for the Multi Frame Lambda
Scan Application.
The advantage of spacing all measurements equally is that presenting
results through use of a spreadsheet is greatly simplified. The operation
returns one wavelength array and a power array for each power meter
channel.
The disadvantage of using equally spaced datapoints is that the linear
interpolation is analogous to the use of a low pass filter. Figure 18 shows
the original curve as measured directly by a Power Meter and the
interpolated curve.
Interpolation will always tend to produce a smoother curve by rounding off
any peaks in the curve.
Original
Curve
Interpolated
Curve

Figure 18 Equally Spaced Datapoints

236
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition