106
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:DISPlay
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:DISPlay
Takes the input power level value as the reference value.
none
none
sens1:pow:ref:disp
All power meters
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe<wsp><boolean>
Sets the measurement units to relative or absolute units.
A boolean value:
none
sens1:pow:ref:stat 1
All power meters
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe?
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe?
Inquires whether the current measurement units are relative (dB) or absolute (Watts or
dBm).
none
A boolean value:
sens1:pow:ref:stat? → 1<END>
All power meters
Master and slave channels are independent.
Measurement Functions – The SENSe Subsystem
0 or OFF: absolute
1 or ON: relative
0: absolute
1: relative
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition