Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
Returns the range setting for the module. For a return loss module, returns the power range
of the return loss diode.
none
The range setting as a float value in dBm
(−110 ≤ value ≤ +30).
sens1:pow:rang?
-2.00000000E+001<END>
All power meters and return loss modules.
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:MONitor[:UPPer]<wsp><value>[DBM]
Sets the power range for a retun loss module's monitor diode.
The range changes at 10 dBm intervals. The corresponding ranges for linear measurements
(measurements in Watts) is given below:
Range
Upper Linear
Power Limit
+30 dBm
1999.9 mW
+20 dBm
199.99 mW
+10 dBm
19.999 mW
1999.9 μW
0 dBm
−10 dBm
199.99 μW
−20 dBm
19.999 μW
−30 dBm
1999.9 nW
−40 dBm
199.99 nW
The range as a float value in dBm. The number is rounded to the closest multiple of 10, be-
cause the range changes at 10 dBm intervals. Units are in dBm.
none
sens1:pow:rang:mon -20DBM
All return loss modules.
Master and slave channels are independent.
Measurement Operations & Settings
Range
Upper Linear
Power Limit
−50 dBm
19.999 nW
−60 dBm
1999.9 pW
−70 dBm
199.99 pW
−80 dBm
19.999 pW
−90 dBm
1.999 pW
−100 dBm
0.199 pW
−110 dBm
0.019 pW
103