GPIB Command Compatibility List

Table 14 Queries with Different Result Values

252

Changed Query Result Values

Table 14 details queries that respond with different return codes than the
old instruments.
Command
*IDN?
*OPT?
*TST
SENS:POW:UNIT?
SOUR:POW:WAV?
SYSTem:ERRor?
SOURce:AM:SOURce?
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition

Changed Query Result Values

Change
Returns new instrument and module identifi-
ers.
Returns new module options.
Selftest result codes are completely new.
0 still means passed.
A head adapter is not overwritten with the
head when it is inserted.
Returns W|DBM not a number.
Returns LOW|UPP|BOTH|EXT and not the
wavelength; use SOUR:WAV? to query the
wavelength. SOUR:WAV:FIXED1? returns the
wavelength of the first laser and
SOUR:WAV:FIXED2? returns the wavelength
of the second laser. For the HP 8153A,
SOUR:POW:WAV? returned the wavelength
of the active laser.
Same functionality but different numbers
and errors are returned for instrument spe-
cific errors.
Returns different enum values than the
HP 8167/8.
Affects
Both
Both
Both
8153
8153
8153
8153
8167/8