:INPut[n][:CHANnel[m]]:OFFSet:POWermeter
:INPut[n][:CHANnel[m]]:OFFSet:POWermeter<wsp><slot>,<channel>
Sets the offset factor (α
powermeter (hosted by the same mainframe) (P
attenuator module's monitor diode (P
α
(dB) = P
(dBm) - P
Offset
att
<slot>
Slot number of the external powermeter.
<channel>
Channel number of the external powermeter.
none
INP1:OFFS:POW 4,2
Attenuator modules with power control.
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed<wsp><value> | MIN | MAX | DEF
Sets the filter transition speed; the speed at which the module moves from one attenuation to
another (in dBs).
<value>
The filter transition speed in dB/s.
MIN | MAX | DEF
Sets the filter transition speed to the module limits, or the module de-
fault.
none
INP1:ATT:SPE 2
All attenuator modules.
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed?
:INPut[n][:CHANnel[m]]:ATTenuation:SPEed?<wsp> MIN | MAX | DEF
Without the optional parameter, queries the transition speed of the filter.
MIN | MAX | DEF Queries the transition speed limits, or the module default.
4 byte Intel floating point; transition speed in dB/s.
INP1:ATT:SPE?
2<END>
All attenuator modules.
) to the difference between a power value measured by another
Offset
ext
).
att
(dBm)
ext
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition